详细信息

The glutamate biosensor and its application to flow injection analysis system.    

文献类型:期刊文献

英文题名:The glutamate biosensor and its application to flow injection analysis system.

作者:Ye, B[1];Li, Q;Li, Y;Yu, J

机构:[1]Research Institute of Biochemical Engineering, East China University of Science and Technology, Shanghai.

年份:1994

卷号:10

期号:2

外文期刊名:Chinese journal of biotechnology

收录:MEDLINE(收录号:7803693);;PubMed

语种:英文

摘要:A micro-enzyme electrode was fabricated by cross-linking L-glutamate oxidase with glutaraldehyde on aminopropyl-platinized platinum wire. A flow injection analysis system with glutamate sensor was used for L-glutamate determination. The peak current is linearly related to the L-glutamate concentration in the range of 0.02-2.0 mM, with good performance, accuracy (CV = 0.4%), fast response (< 60s), and stability (> 20 days). The system was applied to determine the concentration of L-glutamate in a fermentation broth. The recovery rate was in the range of 98.7-107.5%.